Laser light source Ultraviolet / Visible light
LD lasers:

405nm(22mW), 473nm(15mW), 559nm(18mW), 635nm(12mW)

Modulation

Continuously Variable by the LD direct modulation(0.1%-100%, 0.1% inclement)
Line return period - laser OFF

Scanning Scanning method

2 galvanometer scanning mirrors

Scanning mode

Pixel size: 256 × 2564 - 1024 × 1024
Scanning speed: 1.1 s/frame (for pixel size 512 × 512, High Speed scanning mode)
Focusing scanning: High frame rate scan by Y- direction interlace scanning (×1, ×2, ×4)
Dimension: XYT, XYZ, XYZT
Rotation scanning: 0-359.9° in 0.1° increments

Detection Detector module

Fluorescence: 2 channels, Phase Contrast: 1 channel
Variable barrier filter mechanism for fluorescence channel by diffraction grating and slit

Detection method

Analog integration detection by Photomultiplier

Pinhole

Single motorized pinhole
Pinhole diameter: ø50-800μm automatic setting (adjustable to ×1.0, ×1.5, ×2.0, and ×2.5)

Field number

18

Optical zoom

10× objectives: 1× – 6× in 0.1x increments
60× objectives: 1× – 10× in 0.1x increments

Automatic exposure

Automatic setting of the laser intensity and photomultiplier sensitivity to fluorescence intensity.

Focus Z-drive

Motorized focus
Minimum increment: 0.01μm

Objectives

Exclusively designed 10× phase contrast objective / NA 0.4(equivalent to UPLSAPO 10x)
Exclusively designed 60× phase contrast water-immersion objective / NA 1.2 (equivalent to UPLSAPO 60× W)/ with motorized correction collar (•)
Exclusively designed 60× phase contrast oil-immersion objective / NA 1.35 (equivalent to UPLSAPO 60× O) (⊕)
Remote switching from software by electric revolver

Automatic Focus (AF)

Automatic detection of interface between specimen and cover glass by laser reflection light detection
Automatic detection of cover glass thickness and automatic setting of motorized correction collar (•)

Water supply (•)

Automatic water supply and air cleaning mechanism for 60x Water-immersion objective

Oil supply (ο)

Manual
As supporting mechanism, automatic moving of XY stage to oil supply position when switching to 60x

Automatic exposure

Automatic setting of the laser intensity and photomultiplier sensitivity to fluorescence intensity.

XY stage XY driving method

Motorized XY stage module by stepping motor
Minimum increment: 0.3μm

Specimen holder

Only the dedicated specimen holder can be mounted
For three glass bottom dishes with 35mm diameter (•)For one set of cover glass chamber (8 wells type)(•)
Culture pod (for a glass bottom dish with 35mm diameter)(•). For a glass slide
For a glass bottom dish with 35mm diameter ()

Incubator (•) Room environment

Temperature 37±1°C Humidity: more than 90%
CO2 concentration: 5% (recommended), 1 – joint fitting (ø2mm) for exterior CO2 adjustor

Heating method

Non-contact heating by resistive heater mounted on frame section

Control device Controller

Dedicated controller PC/AT-compatible
OS: Windows Vista Business, 32 bit (English version), CPU: Intel Core2Duo 3.0GHz
RAM: 2GB × 2, HDD: 320GB × 2, Special PCI-Express I/F board built-in, Media: DVD-Multi drive built-in

LCD monitor

24 inch LCD monitor × 1, WUXGA (1920×1200)

Main software feature Image acquisition mode

Map image, one shot, time-lapse (XYT), Z-stack (XYZ), Z-stack time-lapse (XYZT), multi area time-lapse (Multi Area XYT), multi
area Z-stack time-lapse (Multi Area XYZT)

Specimen setting

Automatic setting for fluorescence channel and laser according to Dye selected from Dye list

Map image acquisition

Automatic selection of map image of 3×3 – 9x9 fields according to 10× objective lens, and manual selection of map acquisition area

Multi area time-lapse

Automatic multi area time-lapse by motorized XY stage
Setting for each registered point: Image size, scanning speed, cross talk reduction, pinhole diameter,
rotation angle, galvano zoom, acquisition channel, laser power, PMT sensitivity, Z condition
Maximum resister number: 10 items per one container
Maximum interval time: one hour
Maximum acquisition number of times: 3000 times per one point

Image acquisition area

Area appointment: clipping square area in all area (minimum area: 96x96images)

Image display

Display by channel, overlapping display, image in progress review

Cross talk reduction

Line sequential action (2 channel), or frame sequential action (3 channel and 4 channel)

Acquisition image file type

OLYMPUS image format (OIF)

Image file type available
for viewing

OLYMPUS image format (OIF, OIB), Multi-TIFF format (8/16 bit grey scale, index color, 24/32/48 bit color), JPEG, BMP, TIFF

Image editing

LUT: pseudo color setting, contrast adjustment, Comment: inputting graphic, text, scale etc., image extraction, combination

3D image construction

3D display: AlphaBrend method, Maximum intensity projection method 3D animation display, free orientation of cross section display

Image processing

Various types of image filter: Median, Enhanced Edge, etc.
Calculations: inter-image, arithmetic and logical operation

Image analysis

Area and perimeter measurement, time-lapse measurement, colocalization analysis

Room environment Temperature:

18°-28°C

Humidity

30-80% (non condensing)

(•) Applicable only to FV10i-W, (⊕) Applicable only to FV10i-O